Skip to content

Public FDA record

Z-2152-2024: HRIS Storage & Sterilization Case and Tray, Catalog/Part Number: 6210-9-900

Recall record Z-2152-2024 for Howmedica Osteonics Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2152-2024
Company in FDA record
Howmedica Osteonics Corp.
FDA classification date
Jun 18, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The design of the HRIS Storage & Sterilization Case and Tray P/N 6210-9-900 and the Gray Revision Instrument System Accessory Tray/Case P/N 6210-9-910 may lead to misuse of the Gray Revision Instruments including re-use and re-sterilization of the single-use, sterile-packaged instruments.

Source facts

Action
On May 23, 2024 MEDICAL DEVICE RECALL letters were sent to customers. Actions needed: Our records indicate that you may have received the affected product(s). It is Stryker s responsibility as the manufacturer to ensure that customers who may have received these affected products also receive this important communication. We therefore request that you read this notice carefully and complete the following actions. 1. Please inform users of this Medical Device Recall and forward this notice to all individuals who need to be made aware. 2. Immediately check all stock areas and/or operating room storage at your facility to determine if any Gray Revision Instruments from the affected product list in Table 2 have been removed from sterile packaging (Gray Revision Instruments are single-use, sterile-packaged manual surgical instruments). 3. If any Gray Revision Instruments have been removed from sterile packaging, discard/dispose of those instruments. 4. Gray Revision Instruments found within their sterile packaging may be used as intended (single-use, sterile-packaged). 5. Do not use the HRIS Storage & Sterilization Case and Tray or the Gray Revision Instrument System Accessory Tray/Case to re-process (re-sterilize, clean, or re-use) the Gray Revision Instruments, which are single-use, sterile-packaged manual surgical instruments. 6. You may continue to use the HRIS Storage & Sterilization Case and Tray or the Gray Revision Instrument System Accessory Tray/Case for all other instruments as intended. 7. Complete and sign the enclosed Medical Device Recall Business Reply Form and email to strykerortho7963@sedgwick.com/ fax (844) 412-9273. 8. Please contact your Local Sales Office or your Stryker Sales Representative directly for product replacement and inventory questions. Please assist us in meeting our regulatory obligation by emailing back the attached Medical Device Recall Business Reply Form within 5 days. A response is required, even though you may not have any physic
Initiated
2024-05-23
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Brazil, Canada, Chile, Denmark, Hong Kong, Iran, Italy, Japan, Kuwait, Mexico, New Zealand, Singapore, Spain, Sweden, Switzerland, Thailand, Turkey, UK, Venlo.
Recall number
Z-2152-2024
Recall status
Open, Classified
CDRH record ID
208059
FDA classified
2024-06-18
Recall event number
94708

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗