Z-2153-2008: Acu-Loc Distal Radius Plate, Standard, Left. Product labeled in part, " Acu-Loc Distal Radius Plate, Stan

Recall record Z-2153-2008 for Acumed LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2153-2008
Company in FDA record
Acumed LLC
FDA classification date
Aug 28, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Plates are incorrectly laser marked. The PL-DR50L plates are incorrectly laser marked with PL-DR70L.

Source facts

Action
The firm issued an Urgent Notification: Product Removal letter on 6-29-2007. The notification was reportedly sent by either FAX or e-mail. The notification informed consignees to remove the plates from their inventory and return them to Acumed. The notification did not include any instructions for a sub-recall. On 6/13/08, the firm sent a second Urgent Notification: Product Removal letter. This letter describes the product as PL-DR50L - Lot Code - W48003, explains the problem and asks customers to return product to Acumed.
Initiated
2007-06-29
Root cause
Process control
Terminated
2010-02-03
Distribution
Worldwide Distribution --- including USA and countries of South Africa, Italy, Finland, and Brazil.
Recall number
Z-2153-2008
Recall status
Terminated
CDRH record ID
71481
FDA classified
2008-08-28
Recall event number
48466

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗