A software anomaly has been identified with the Import Export Application version 11.0 where
MLC leaf positions for plans using BrainLab m3 and mMLC are incorrectly imported.
Source facts
Action
Varian Medical Systems sent a " Urgent Medical Device Correction Urgent Field Safety Notice" dated July 2, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
Customers were instructed until a correction is deployed all treatment plans using the BrainLab m3 mMLC must be imported using the Import Wizard only.
Varian Medical Systems is developing a technical correction for this issue. You will be contacted by a Customer Service Representative when this correction is available to schedule its installation on your system.
For further questions please call 1.888.827.4265.
Initiated
2012-07-02
Root cause
Software design
Terminated
2013-07-01
Distribution
Nationwide Distribution including the states of New York and Ohio.
Recall number
Z-2153-2012
Recall status
Terminated
CDRH record ID
111025
FDA classified
2012-08-08
Recall event number
62562