Z-2153-2017: Opteryx Variable Rescue Screw, Part #CP2214, UDI: *+M711CP22140/$$801716012*. Packaged in heat-sealed pl

Recall record Z-2153-2017 for Innovasis, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2153-2017
Company in FDA record
Innovasis, Inc
FDA classification date
May 23, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Opteryx Variable Rescue Screws are engraved with incorrect (larger) size (16 instead of 14).

Source facts

Action
Distributors are being notified by e-mail, with attached copies of the written letter and forms to be printed, signed and returned to Innovasis on 5/5/17. Additional follow-up may be done via telephone and fax to return products.
Initiated
2017-05-05
Root cause
Error in labeling
Terminated
2017-08-01
Distribution
US Nationwide Distribution in the states of CA, UT, ID, TX, OH, and MN.
Recall number
Z-2153-2017
Recall status
Terminated
CDRH record ID
155431
FDA classified
2017-05-23
Recall event number
77206

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗