Z-2153-2023: Simplexa Direct Amplification Disc Kit, Rx Only, IVD The Simplexa Direct Amplification Disc Kit, REF: MO

Recall record Z-2153-2023 for DiaSorin Molecular LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2153-2023
Company in FDA record
DiaSorin Molecular LLC
FDA classification date
Jul 18, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for the direct amplification disc to malfunction which may result in spillage of liquid from the disc.

Source facts

Action
On June 19, 2023 DiaSorin Molecular issued a "urgent: Medical Device Correction Notification to affected consignees via E-Mail. DiaSorin Molecular asked consignees to take the following actions: 1. Any remaining discs in inventory that are listed in the table below should be discarded. Replacement will be provided. 2. Any sample with an error code should be retested. 3. If leakage is suspected, the spillage should be removed, and the instrument surface should be decontaminated. Ensure the practice of Good Laboratory Procedures including regular cleaning. 4. Notify others within your network that may have received these kits. 5. Complete and return the Acknowledgement and Receipt Form (page 3) of this notification letter. 6. If your firm has further distributed the devices, please share a copy of this correction notification and response form with those customers and provide a copy of the completed response form back to us. 7. If you have any questions, and to request material replacement, please contact DiaSorin Molecular Technical Service, Monday Friday, 7:00am to 5:00pm (Pacific Time) at (800) 838-4548, option 3, or by email at technicalinfo.molecular@DiaSorin.com. 8. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax
Initiated
2023-06-14
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SD, TN, TX, VA, VT, WA, WI, WV and the countries of Bangladesh, Canada, Israel, Italy.
Recall number
Z-2153-2023
Recall status
Open, Classified
CDRH record ID
201596
FDA classified
2023-07-18
Recall event number
92559

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗