Skip to content

Public FDA record

Z-2154-2008: Corometrics MODEL 250 Series Maternal/ Fetal Monitor, Models 256 and 259, part number 2024489-002. Inten

Recall record Z-2154-2008 for Datex Ohmeda, Inc. dba GE Healthcare. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2154-2008
Company in FDA record
Datex Ohmeda, Inc. dba GE Healthcare
FDA classification date
Aug 26, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Communication connector within monitor is defective and may cause potential loss of telemetry.

Source facts

Action
The recalling firm notified healthcare professionals by an Urgent Medical Device Correction letter dated 07/01/2008 to Healthcare Administrators, Risk Managers, Nurse Managers and Directors of Clinical Engineering. The letter advised of a defect in a communication board that may cause a potential loss of telemetry monitoring. The firm recommended that the device should not be used with the firm's 340 Series telemetry system pending replacement of the defective board by service engineers. The Healthcare Professionals are asked to contact the GE Healthcare Technical Support center at 1-800-558-7044.
Initiated
2008-05-05
Root cause
Component design/selection
Terminated
2012-08-07
Distribution
Nationwide Distribution.
Recall number
Z-2154-2008
Recall status
Terminated
CDRH record ID
71483
FDA classified
2008-08-26
Recall event number
48467

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗