Communication connector within monitor is defective and may cause potential loss of telemetry.
Source facts
Action
The recalling firm notified healthcare professionals by an Urgent Medical Device Correction letter dated 07/01/2008 to Healthcare Administrators, Risk Managers, Nurse Managers and Directors of Clinical Engineering. The letter advised of a defect in a communication board that may cause a potential loss of telemetry monitoring. The firm recommended that the device should not be used with the firm's 340 Series telemetry system pending replacement of the defective board by service engineers. The Healthcare Professionals are asked to contact the GE Healthcare Technical Support center at 1-800-558-7044.
Initiated
2008-05-05
Root cause
Component design/selection
Terminated
2012-08-07
Distribution
Nationwide Distribution.
Recall number
Z-2154-2008
Recall status
Terminated
CDRH record ID
71483
FDA classified
2008-08-26
Recall event number
48467
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.