Skip to content

Public FDA record

Z-2154-2012: 3mL 9NC Coagulation Sodium Citrate 3.2%, Item #454334, Greiner Bio-One North America Inc., 4238 Capital

Recall record Z-2154-2012 for Greiner Bio-One North America, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2154-2012
Company in FDA record
Greiner Bio-One North America, Inc.
FDA classification date
Aug 7, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

Part of the lot received a lower concentration of citrate solution which may cause falsely lower values for common coagulation tests.

Source facts

Action
Distributors were notified by letter on 07/02/2012 and customers were sent a letter dated 07/05/2012. The letters identified the affected product and the reason for the recall. Distributors were asked to stop distributing the affected product and to complete the attached Product Disposition form. The form along with the affected product was to be returned to GBO. The firm will provide replacement stock or issue a credit once product return and receipt of the Product Disposition form has been confirmed. Customers were also asked to complete the Product Disposition Site Confirmation form and fax it back to Greiner Bio-One North America, Inc. at the number provided. The letter states that the firm will replace any affected product once the form is received. Questions should be directed to 800-515-8112.
Initiated
2012-07-02
Root cause
Nonconforming Material/Component
Terminated
2013-06-11
Distribution
Nationwide Distribution
Recall number
Z-2154-2012
Recall status
Terminated
CDRH record ID
111118
FDA classified
2012-08-07
Recall event number
62595

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 21, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗