Skip to content

Public FDA record

Z-2154-2017: Philips DigitalDiagnost 3.1.x X-Ray System

Recall record Z-2154-2017 for Philips Electronics North America Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2154-2017
Company in FDA record
Philips Electronics North America Corporation
FDA record creation date
Jun 9, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During recent evaluations of the Philips DigitalDiagnost 3.1.x, we have identified a potential issue that may affect the performance of the equipment under certain conditions.

Source facts

Action
Philip's planned action to bring defect into compliance: 1. You will contact customers and initiate a software update to correct the issue that incorrectly notifies the user of the detectors being ready. 2. You will install the software/firmware and perform testing to ensure the software update was effective. 3.You will provide a customer notification letter that includes a statement that you will without charge, remedy the defect or bring the product into compliance. CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2.Philips Healthcare will implement this CAP by November 10, 2017. If you need any further information or support concerning this Issue, please contact your local Philips representative: Technical Support Line: 1-800-722-9377. We apologize for any inconvenience this may cause and trust that this information is adequately addressing any concerns you may have.
Initiated
2017-05-11
Root cause
Software design
Terminated
2018-07-10
Distribution
USA(nationwide) Distribution to the states of : AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KY, LA, MA, ME, MI, MN, MO, NC, ND, NE, NH, NM, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA and WI
Recall number
Z-2154-2017
Recall status
Terminated
CDRH record ID
155677
FDA record created
2017-06-09
Recall event number
77312

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗