Open original FDA record ↗Recall record
- FDA identifier
- Z-2154-2019
- Company in FDA record
- SenTec AG
- FDA record creation date
- Aug 5, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is an increased chance (approx. 5%) of no electrolyte or a reduced amount of electrolyte being applied to the sensor during the membrane change procedure.
Source facts
- Action
- On May 2, 2019, the firm sent Guidance for SenTec Distributors to their distributors advising them of the issue, requesting them to check their stock and notify their end users via a voluntarily initiated field correction action associated with Membrane Changers and their Inserts which was enclosed with the "recall" letters. The distributors were to report back to SenTec via email.
- Initiated
- 2019-05-02
- Root cause
- Process control
- Terminated
- 2020-12-16
- Distribution
- Worldwide distribution - US Nationwide in the state of Missouri and countries of Argentina, Australia, Austria, Belgium, Chile, China, Columbia, Costa Rica, Czech Republic, Denmark, Ecuador, Europe, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Indonesia, Israel, Italy, Kuwait, Malaysia, Finland, Portugal, Romania, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, United Arab Emirates and United Kingdom.
- Recall number
- Z-2154-2019
- Recall status
- Terminated
- CDRH record ID
- 174652
- FDA record created
- 2019-08-05
- Recall event number
- 83115