Open original FDA record ↗Recall record
- FDA identifier
- Z-2155-2014
- Company in FDA record
- Cardiovascular Systems, Inc.
- FDA classification date
- Aug 14, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
CSI has initiated a recall on the Diamondback 360 Peripheral Orbital Atherectomy Device because it may contain a saline sheath that may experience cracking, fracture, and release particulate during use.
Source facts
- Action
- Cardiovascular System, Inc. sent consignees an Urgent Medical Device Recall letter dated May 23, 2014. The letter described the Affected Product, Recall Description, Instructions which included to remove the affected product and return it to CSI and to complete and return the Customer Acknowledgement Form. For further Information they customers were instructed to contact Customer Service, Cardiovascular System, Inc., 877-274-0901. For questions regarding this recall call 877-274-0901.
- Initiated
- 2014-05-23
- Root cause
- Process control
- Terminated
- 2014-10-01
- Distribution
- Nationwide Distribution including AZ, AR, CA, CT, FL, IL, IA, MD, MI, NY, NC, PA, TN,and TX.
- Recall number
- Z-2155-2014
- Recall status
- Terminated
- CDRH record ID
- 128772
- FDA classified
- 2014-08-14
- Recall event number
- 68806