B. Braun is voluntarily recalling 13 lots of CytoGuard Closed Luer Connector due to a potential for some blisters to be punctured resulting in a compromised sterility barrier.
Source facts
Action
B. Braun mailed an Urgent Medical Device Recall Notice to affected customers to inform them of the issue. Customers were asked to determine their current inventory and to not destroy any of the product; however, product should be discontinued immediately and quarantined and return the acknowledgement form for product return.
Initiated
2017-04-10
Root cause
Nonconforming Material/Component
Terminated
2018-02-02
Distribution
United States Nationwide distribution
Recall number
Z-2155-2017
Recall status
Terminated
CDRH record ID
155027
FDA record created
2017-05-23
Recall event number
77112