Z-2155-2021: Proteus 235-Proton Therapy System

Recall record Z-2155-2021 for Ion Beam Applications S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2155-2021
Company in FDA record
Ion Beam Applications S.A.
FDA record creation date
Jul 28, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

if the user forgets to select the trigger input manually while beam gating is prescribed for the patient, the proton irradiation will be delivered without gating. If the beam is not gated as prescribed, the treatment field may be delivered with an error in position

Source facts

Action
IBA issued Urgent Medical Device Correction notification starting on May 27, 2021. Letter states reason for recall, health risk and action to take: IBA would like to remind Proteus235 users of the following warning present in the Clinical User Guide: As a Radiation Therapy Technologist (RTT), check before allowing the irradiation that the position of the selector switch of the external beam triggering interface is consistent with the prescription data in the patient file. There is a risk of mistreatment as the system does not automatically detect the gating prescription thereby always generating a continuous irradiation. Labelling change and training IBA will: - update the user manuals to recommend the customers to define a procedure for patients with gating prescribed in order to ensure that the system will be set correctly, - provide user training sessions about the preparation of a treatment with UBTI synchronization. The actions will be performed for your site by December 31, 2021. Your IBA representative is able to provide you with additional information and/or guidelines if necessary.
Initiated
2021-05-26
Root cause
Labeling design
Distribution
FL, PA, TN, TX, VA Foreign: Germany, Italy, Russia, Sweden , India, Netherlands
Recall number
Z-2155-2021
Recall status
Open, Classified
CDRH record ID
188066
FDA record created
2021-07-28
Recall event number
88196

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗