An extremely dusty computer can cause problems at system start, or rarely, cause system fail. Perform a system check prior to performing exams. If the Sensis system fails, its functions can not be used. Clinical treatment may need to be terminated, restarted, or transferred to a functioning system.
Source facts
Action
Siemens mailed a Customer Safety Advisory Notice (CSAN) to affected customers on 04/12/2017 to inform them of the issue. The Notice explained the issue and when it can occur, the possible risks involved, and the actions consignees should take to address the issue. For questions regarding this recall call 610-219-4834.
Initiated
2017-04-12
Root cause
Material/Component Contamination
Terminated
2018-05-15
Distribution
Nationwide Distribution
Recall number
Z-2156-2017
Recall status
Terminated
CDRH record ID
154828
FDA record created
2017-05-24
Recall event number
77016