Z-2156-2021: Affinity Four Birthing Bed-used as a birthing bed for women of child bearing age in an LDR (Labor, Delive

Recall record Z-2156-2021 for Hill-Rom, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2156-2021
Company in FDA record
Hill-Rom, Inc.
FDA record creation date
Jul 28, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Latch mechanism on the Lift-Off foot section of the Affinity Four Birthing Bed is damaged, it could potentially cause the installed foot section to be improperly engaged onto the bed, could result in injury to the user due to a fall.

Source facts

Action
Hill Rom Issued Urgent Medical Device Correction Subject: Affinity Four Birthing BedLift-Off Foot Section REF: MOD1331 issued on 05/28/21 to Facility Risk Manager; Facility Administrator; Facility Engineer; Vigilance Manager; Biomedical Engineering; Medical Device Liaison Officer; Distributor, stating reason for recall, health risk and action to take: Verify that the Lift-Off foot section has already been updated with the new Latch Bracket and Foot Section receivers and has adequate engagement into the receivers as noted in the graphics below. Be aware that the foot section has the potential to fall if the latch brackets and foot section receivers have not been updated and are damaged. Refer to Figures 2 and 3 below and complete the following inspection process to determine if your Latch Bracket and Foot Section receiver have adequate engagement. Action to be taken by the distributor: Please share with your end users to instruct them on completing their inspection test outlined above. Complete the attached response form and return to hillrom1331@stericycle.com without delay. Contact hillrom1331@stericycle.com to receive an electronic copy of this notification and response form for onward distribution. Hillrom will mitigate this risk through an update to the Lift-Off foot section and will contact you to coordinate the update. Once the product has been updated, some customers who utilize the P278 Hillrom Instrument Tray, may encounter issues fitting the tray. These original version trays will be replaced, as needed, at the time of the update. If you have any questions regarding this safety notice, please contact Hillrom Technical Support at 800-445-3720 or us.service@hillrom.com.
Initiated
2021-05-28
Root cause
Device Design
Terminated
2024-06-06
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, CA, IA, MS, NC, NY, OH, PA, TX, VA and the countries of Australia, Austria, Azerbaijan, Belgium, Chile, China, Denmark, Finland, France, Germany, Ireland, Israel, Italy, Jordan, Kuwait, Netherlands, New Zealand, Norway, Portugal, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom.
Recall number
Z-2156-2021
Recall status
Terminated
CDRH record ID
188056
FDA record created
2021-07-28
Recall event number
88185

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗