Affected convenience kits contain BD ChloraPrep Clear - 1 mL Applicators, which were recalled by BD due to potential to exhibit an open seal on its packaging. This may constitute a breach to sterility, which may in turn lead to the infection related harms for the patient.
Source facts
Action
On June 12, 2025, firm instructed customers to request stickers to over-label the affected convenience kits with a warning label indicating to remove the affected component and replace it with product from supply. No product is to be returned.
Initiated
2025-06-12
Root cause
Other
Distribution
Distributed domestically to California, Florida, Pennsylvania, and Texas. No international distribution.
Recall number
Z-2156-2025
Recall status
Open, Classified
CDRH record ID
214766
FDA classified
2025-07-21
Recall event number
97154
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.