Open original FDA record ↗Recall record
- FDA identifier
- Z-2158-2012
- Company in FDA record
- Biomet 3i, LLC
- FDA classification date
- Aug 8, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Biomet 3i is recalling their product Full Osseotite Tapered Certain Implant IFNT611. The depth of the implant's internal hex is too shallow. The condition, if present, may prevent the driver; cover screw or abutment from fully engaging the implant.
Source facts
- Action
- BIOMET 3i sent an Urgent Medical Device Recall letter on March 19, 2012, to all affected customers via email or fax. The letter identified the product, the problem, and the action to be taken by the customer. Customers are to return any unopened and unused product for replacement at no cost. In addition, customers should contact Customer Service at 1-800-342-5454 for assistance if they encounter issues with the mating component, or if they have any questions or concerns. Customers are to respond to the notification regardless of whether they have product to return.
- Initiated
- 2012-03-23
- Root cause
- Process control
- Terminated
- 2013-12-16
- Distribution
- Worldwide Distribution -- United States, UK, France, Canada, Mexico, Poland, Lebanon and Hong Kong.
- Recall number
- Z-2158-2012
- Recall status
- Terminated
- CDRH record ID
- 110364
- FDA classified
- 2012-08-08
- Recall event number
- 62375