Z-2158-2021: Zippie Voyage, early intervention stroller - Product Usage: is manually operated by a caregiver for purpo

Recall record Z-2158-2021 for Sunrise Medical (US) LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2158-2021
Company in FDA record
Sunrise Medical (US) LLC
FDA record creation date
Jul 28, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The seating system unexpectedly detached, which resulted in the seating system falling to the ground, causing injury.

Source facts

Action
On 07/12/2021, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Notification informing customers that if the seating system is not adequately attached to the stroller base, the seating system may slide off the base and fall to the ground potentially causing injury a child. The Recalling Firm will be providing a safety tether and are informing/instructing their customers to: This Safety Tether Kit will be provided free of charge to all consumers starting the week of July 12, 2021. The Safety Tether Kit will be accompanied by instructions to guide consumers through installing and using the safety tether. The attached consumer communication template letter provide the consumer with three options until their Safety Tether Kit is received and the safety tether has been installed by them: 1. Choose to stop using the ZIPPIE Voyage early intervention stroller until a Safety Tether Kit has been provided and the safety tether has been installed by the consumer. OR 2. If they can elect to continue to use the early intervention stroller (before receiving their Safety Tether Kit and having installed the safety tether) and they can order their stroller with a transit option, they can use the transit strap to securely attach the stroller seat to the stroller handle. The transit strap may function to prevent the seat, if inadequately attached to the stroller base, from falling to the ground. However, the transit strap is not intended to be a replacement for the Safety Tether Kit that we will be providing. The tether set should be installed and used once it is received. OR 3. Experienced users who do not have the transit option on their stroller and choose to continue to use the stroller (before they receive their Safety Tether Kit and have installed the safety tether), must ensure that the stroller seat is securely attached to the stroller base before every use (following the instructions shown in the Owners Manual]). These steps must be performed each and every time the seat
Initiated
2021-07-08
Root cause
Device Design
Terminated
2025-03-28
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of United Kingdom, Germany, France, Canada, Italy, Netherlands, Poland, Czech Republic, Australia, Hong Kong, Japan, Qatar, Norway, United Arab Emirates, China, Saudi Arabia, Sweden, Switzerland, Spain, Indonesia, India, Singapore, Bahrain and Israel.
Recall number
Z-2158-2021
Recall status
Terminated
CDRH record ID
188302
FDA record created
2021-07-28
Recall event number
88257

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗