Skip to content

Public FDA record

Z-2159-2008: 10% Buffered Formalin Phosphate; Certified For laboratory and manufacturing use only, not for drug, food

Recall record Z-2159-2008 for ThermoFisher Scientific Co., LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2159-2008
Company in FDA record
ThermoFisher Scientific Co., LLC
FDA classification date
Aug 25, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Assay tests on retention sample of 10% Buffered Formalin Phosphate showed it to be 3.417%, which is below the specification of 4-5%.

Source facts

Action
Urgent: Product Recall notification letters were sent by US mail on May 1, 2008. The letters asked that those in receipt of the product check their stock for this product, destroy any of the material that remains from the referenced lot, and then contact their local customer service center for a replacement. The letter also states that the appropriate corrective measures have already been instituted to prevent a recurrence.
Initiated
2008-05-01
Root cause
Nonconforming Material/Component
Terminated
2012-08-13
Distribution
Worldwide Distribution --- including USA and Canada.
Recall number
Z-2159-2008
Recall status
Terminated
CDRH record ID
71704
FDA classified
2008-08-25
Recall event number
48601

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗