Skip to content

Public FDA record

Z-2159-2019: DRG Salivary Testosterone ELISA- IVD for the detection of Testerone in human saliva Catalog # SLV-3013

Recall record Z-2159-2019 for DRG Instruments GmbH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2159-2019
Company in FDA record
DRG Instruments GmbH
FDA record creation date
Aug 5, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Salivary assay may observe a higher percentage of samples with 0 pg/ml testosterone or results which are too low. Saliva samples in the low measuring range of the kit(< 50 pg/ml) were assayed significantly lower

Source facts

Action
DRG International issued notification via email on 5/21/19 advising of the problem, health risk and action to take: 1. We kindly ask you to immediately examine your stock and promptly quarantine these products, 2. Please send back or discard all unused kits. 3. Please use the attached response form, fill, sign and send back to us.4. Please inform your customers and forward this letter and notice to persons who received the affected lots and request they fill out the form and return at once.5. Patient samples that have been determined with the affected lots should be checked again carefully. If results are lower than expected samples should be re-run again. 6. Please tell us if you have experienced any problem with this lot and specify the details. Reply to DRG (email: qa@drg-international.com and copy drqsupport@drginternational.com).
Initiated
2019-05-21
Root cause
Under Investigation by firm
Terminated
2020-05-29
Distribution
Worldwide Distribution - US Nationwide CA, MI Foreign: Canada, Brazil, China, Poland
Recall number
Z-2159-2019
Recall status
Terminated
CDRH record ID
173940
FDA record created
2019-08-05
Recall event number
83183

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 21, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗