Z-2159-2024: Inspire Model 3028, IV Implantable Pulse Generator

Recall record Z-2159-2024 for Inspire Medical Systems Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2159-2024
Company in FDA record
Inspire Medical Systems Inc.
FDA classification date
Jun 21, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

There is a potential manufacturing defect which can lead to electrical leakage in the sensing circuit which causes a system malfunction and requires a revision surgery to replace the IPG,

Source facts

Action
Inspire Medical systems issued an URGENT Medical Device Correction notice to the affected healthcare providers on 05/16/2024 via FedEx overnight. The notice explains the potential problem with the device and risk to the patient and recommends the following actions: We recommend the following actions: " Notify affected patients of this voluntary recall. " Schedule your patient for a monitoring visit to confirm their Inspire therapy is working as intended through waveform and impedance analysis. Monitor patients for reports of variation in stimulation, a lack of therapy effectiveness, or an inability to turn on therapy. " Continue with routine monitoring and conduct waveform and impedance analysis at every routine office visit, as IPG issue may be non-invasively identified with basic waveform and impedance testing. " If abnormal waveforms or impedances are observed, record impedance values at 1.5V, 2.0V, 2.5V, 3.0V and 3.5V. o Let the patient know they will feel 3 short pulses of stimulation. o Some of these stimulation pulses may feel strong. This is normal. " Contact your Inspire representative for additional diagnostic testing and troubleshooting recommendations. For additional questions: please contact your Inspire representative or the Inspire Quality team at ProductRecalls@inspiresleep.com.
Initiated
2024-05-16
Root cause
Process control
Distribution
Worldwide distribution - US Nationwide and the country of Germany.
Recall number
Z-2159-2024
Recall status
Open, Classified
CDRH record ID
207939
FDA classified
2024-06-21
Recall event number
94658

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗