Z-2160-2021: OEC 3D Imaging, Image-Intensified Fluoroscopic X-Ray System, Mobile - Product Usage: designed to provide

Recall record Z-2160-2021 for GE OEC Medical Systems, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2160-2021
Company in FDA record
GE OEC Medical Systems, Inc
FDA record creation date
Jul 29, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential that the coin cell battery used to monitor X-Ray tube temperature may deplete early causing the system to become inoperable.

Source facts

Action
On June 11, 2021, GE Healthcare issued an "Urgent Medical Device Correction via certified mail to all affected consignees. In addition to informing consignees about the recalled product, the firm asked customers to take the following actions: 1. You may continue to use your system. 2. Please disseminate this information to applicable users within your facility. 3. When the system is not in use, it is recommended to connect the system to power with the workstation plugged in and the interconnect cable attached to the mainframe, to reduce potential battery depletion. It is not necessary to turn the workstation power on. 4. Prior to each use, please follow the Start Up Checklist in the System Set Up section (Section 2) of the OEC Elite or OEC 3D Operator Manuals and verify that no error message is displayed on the touch panels or the monitor. a. If the system displays the following error message, as described in the Troubleshooting and messages section (Section 14) of the OEC Elite or OEC 3D Operator Manuals: , this is an indication of a depleted coin cell battery. If this error is displayed, please contact GE Authorized Service. 5. Complete and return the attached response form promptly upon receipt and return to fieldactionssurgery@ge.com. 6. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative. 7. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Initiated
2021-06-10
Root cause
Under Investigation by firm
Terminated
2024-05-09
Distribution
Worldwide distribution - US Nationwide distribution including in the states of CA, PA, WA, OH, LA, MS, AL, FL, MO, OK, AZ, NJ, CO, MI, NV, MN, KS, TN, NC, IL, MD, SC,GA, UT, WI, MT, AR, WV, ID,CT, IA, WY, VT, KY, DC, SD,ME, Virgin Islands and the countries of Colombia, Egypt, Peru, Chile, Spain, France, Germany, Hong Kong, Canada, Japan, Brazil, Turkey, United Arab Emirates, Singapore, Australia, Thailand, United Kingdom, Italy, Korea, Russia, CHINA, Romania, Nigeria, Czechia, Argentina, Taiwan, Bangladeh, Malaysia, Bosnia and Herzegovina, Poland, Belgium, Indonesia, Dominican Republic, Switzerland, Hungary, New Zealand, Vietnam, Mexico, India, Norway, Philippines, Ecuador, Lithuania, HONDURAS, Uruguay, Sweden, Croatia, Jamaica, Kuwait, Ethiopia, Saudi Arabia, Finland, Bahamas, Trinidad and Tobago, Pakistan, Slovakia, Costa Rica, Israel, Panama, Greece, South Africa and DENMARK.
Recall number
Z-2160-2021
Recall status
Terminated
CDRH record ID
188125
FDA record created
2021-07-29
Recall event number
88202

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗