Open original FDA record ↗Recall record
- FDA identifier
- Z-2161-2018
- Company in FDA record
- Maquet Cardiopulmonary Ag
- FDA record creation date
- Jun 11, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
This lot of arterial cannula may have been supplied with 3/8-width de-airing plugs that could break during the process of removal from the arterial cannula.
Source facts
- Action
- Urgent Medical Device Recall letters were distributed beginning 5/14/18. The letters instructed customers to do the following: 1. Read and understand this notification, and disseminate the information provided to all users, or prospective users, of the affected arterial cannulae. 2. Check your arterial cannula inventory and compare with the Getinge part number, article number and lot number provided on Page 1 of this notification. 3. Remove any affected product that are still in your inventory and prepare it for shipment back to Getinge. 4. Call Getinge Customer Support at 1-888-627-8383 (Press Option 2 and then Option 2) for a Return Authorization (RGA) and for shipping information and label. Customer Support will issue a credit for the product return. 5. Complete, sign and promptly return the filled out response form (Page 4 of this notification) either by email to FieldActions@getinge.com or by fax to 1-973-860-1675.
- Initiated
- 2018-05-14
- Root cause
- Component design/selection
- Terminated
- 2020-05-15
- Distribution
- Worldwide distribution including US state of VA and Puerto Rico, Australia, Belgium, Brazil, Canada, Chile, Colombia, Cuba, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Iran, Ireland, Israel, Italy, Japan, Lebanon, Malaysia, Martinique, Mauritius, Mexico, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Reunion, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Syria, Thailand, Turkey, United Arab Emirates, and United Kingdom.
- Recall number
- Z-2161-2018
- Recall status
- Terminated
- CDRH record ID
- 164622
- FDA record created
- 2018-06-11
- Recall event number
- 80076