Z-2161-2026: LUCAS 2 Model Number: 99576 Catalog Numbers: 99576-000024 and 99576-000011 UDI-DI code: 00883873904565 L

Recall record Z-2161-2026 for Jolife AB. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2161-2026
Company in FDA record
Jolife AB
FDA classification date
May 11, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a subset of chest compression systems/devices where the service performed on the systems were not appropriately documented (The documents for the service performed to justify the passing results were not include in the service records).

Source facts

Action
On 04/13/2026, the firm sent via US Postal an "Urgent: Medical device correction" letter informing customers that through complaint reviews, Stryker discovered a subset of work orders performed between July 2023 through November 2025 with Performance Inspection Procedures (PIP) recorded as PASS without the corresponding objective evidence properly documented in the work orders. Customers are instructed to: 1. Immediately check internal inventory to locate the affected product as listed. 2. Return the business reply form by email to RSRecall@Stryker.com to confirm receipt of this notification. 3. Upon receipt of the completed business reply form, Stryker will contact customers to arrange for corrective action on the device if service has not been performed. 4. There are no additional actions that users can and should take. Continue to follow the Operating Instructions and perform the daily checks and user tests. These tests can be found in the chapter Maintaining the Equipment subsection General Maintenance and Testing . 5. Maintain awareness of this communication internally until all required actions have been completed within their facility. Ensure this letter is kept with the affected device until the correction has been completed. 6. Inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details, so Stryker can inform the recipients appropriately. Any questions or concerns, contact Customer Service +1 800 787 9537, option 2, from 8:00 AM to 7:00 PM (Eastern Time), Monday Friday or by email at medtechsup@stryker.com.
Initiated
2026-04-13
Root cause
Under Investigation by firm
Distribution
U.S. Nationwide distribution in the states of AL, AZ, CA, FL, HI, IA, KS, KY, LA, MA, MD, MN, NC, NV, OH, SC, TX, UT, VA, WA, and WY.
Recall number
Z-2161-2026
Recall status
Open, Classified
CDRH record ID
218655
FDA classified
2026-05-11
Recall event number
98417

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗