Skip to content

Public FDA record

Z-2162-2014: Arterial Transpac IT Monitoring Kit w/Safeset Reservoir, 03 ml Flush Device and 2 CSP, Item No. 011-46106

Recall record Z-2162-2014 for ICU Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2162-2014
Company in FDA record
ICU Medical, Inc.
FDA classification date
Aug 7, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

ICU has become aware of an issue with a component where the connection on an arterial monitoring kit may crack and cause a fluid leak.

Source facts

Action
ICU Medical, Inc. sent an email dated July 18, 2014, to their customers, The email inform the customer of the problem and the action to be taken. The email stated that the US Sales Representative would remove the recalled product on July 22, 2014. If product was further distribution customers should notify their customers. Customer was informed of the RGA to have the recalled product returned. For questions regarding this recall call 801-264-1732.
Initiated
2014-07-18
Root cause
Nonconforming Material/Component
Terminated
2014-09-25
Distribution
Worldwide Distribution - USA including Oregon and Internationally to Australia.
Recall number
Z-2162-2014
Recall status
Terminated
CDRH record ID
129168
FDA classified
2014-08-07
Recall event number
68971

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗