Skip to content

Public FDA record

Z-2162-2017: CORFLO, PEG Adaptor Repair Kit with ENFit Connectors, Size: 16FR; Product Usage: CORFLO PEG Tubes are i

Recall record Z-2162-2017 for CORPAK MedSystems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2162-2017
Company in FDA record
CORPAK MedSystems, Inc.
FDA record creation date
May 25, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Firm is initiating voluntary correction of CORFLO PEG with ENFit. Barb connector (stem) of the adapter may break unexpectedly when exposed to stress.

Source facts

Action
Firm provided consignees with field correction on 2/2/2017 by letter mailed via FedEx. Within the notice, the firm provided additional instructions for handling the PEG Y-Adapter to help reduce the likelihood for exposing the stem to excess stress that could break the connector. Consignees were instructed to forward the field correction notice to all staff members or end-users who use the ENFit Y-Adapter.
Initiated
2017-02-03
Root cause
Device Design
Terminated
2019-08-05
Distribution
Worldwide distribution in the states of NH, VT, CA and the countries of UK, Ireland, France, Austria, Norway, Sweden, New Zealand
Recall number
Z-2162-2017
Recall status
Terminated
CDRH record ID
155499
FDA record created
2017-05-25
Recall event number
76465

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗