Z-2162-2020: GENTLEWAVE REF: FG-002-0001

Recall record Z-2162-2020 for Sonendo Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2162-2020
Company in FDA record
Sonendo Inc
FDA record creation date
May 28, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The console would continue to run for extended period when the foot pedal was released.

Source facts

Action
On 10/23/2019, "Urgent: Medical Device Recall" of a foot pedal, the firm provided information on how to recognize the device malfunction as: If the treatment fluid is continuously cycling after the foot pedal is released, this is a sign that the foot pedal is not working properly. The firm instructed customers to replace the current foot pedal with the new foot pedal that was sent with this notification. and instructed to: Attach the new foot pedal to the Foot Pedal connection located at the lower rear of the Console as instructed per IFU. The firm instructed customers to return the current foot pedal in the provided pre-paid FedEx box and send the enclosed acknowledgement letter to the firm by mail. The contact information was provided to customers as: Calls Monday through Friday 5:00 AM to 5:00 PM (Pacific Time) 1-844-468-5928 or email complaints@sonendo.com.
Initiated
2019-10-23
Root cause
Device Design
Terminated
2022-01-20
Distribution
U.S.: OR, CA, NY, SD, CO, AL, WA, IN, TX, NM, IL, NC, AZ, TN, KS, MT, FL, UT, OK, ID, OH, ME, WV, CT, GA, MO, MI, NE, KY, VT, ND, NJ, DC, LA, NY, MA, VA, MN, SC, MD, PA, NH, NV, NE, OR, MS, WI, AK, IA. No foreign consignees.
Recall number
Z-2162-2020
Recall status
Terminated
CDRH record ID
181303
FDA record created
2020-05-28
Recall event number
85593

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗