Skip to content

Public FDA record

Z-2162-2025: BD COR System Software. Model Number: 444829.

Recall record Z-2162-2025 for Becton Dickinson & Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2162-2025
Company in FDA record
Becton Dickinson & Co.
FDA classification date
Jul 23, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for functionality issue that supports the Over labeling feature and could lead to possible false negative results.

Source facts

Action
Becton, Dickinson and Company began notification of consignees on about 06/23/2025 via telephone call and followed by a letter. Consignees were instructed to note the outlined workflows when loading specimens with the Over labeling feature enabled and consider retesting affected specimens for HPV testing if affected. Consignees were also instructed to post the notification within their facility, notify customers if affected units were further distributed, and complete and return the provided Customer Response Form.
Initiated
2025-06-23
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution in the states of Florida, Texas and Michigan. The countries of Sweden and Belgium.
Recall number
Z-2162-2025
Recall status
Open, Classified
CDRH record ID
214649
FDA classified
2025-07-23
Recall event number
97142

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗