B. Braun Medical Inc. has become aware of an issue with the Outlook ES Safety Infusion System. The key panel may fail to function as intended if the underlying circuitry fails due to intermittent contact. This issue applies specifically to the HOLD button. If this occurs, the Door Processor (DP) will not receive the desired input from the user. This issue can be observed in pumps with key panels
Source facts
Action
B Braun sent an URGENT: MEDICAL DEVICE FIELD CORRECTION letter dated October 24, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact Customer Support at 1-800-627-7867 for questions regarding this notice.
Initiated
2012-10-24
Root cause
Component design/selection
Terminated
2015-07-27
Distribution
Nationwide Distribution.
Recall number
Z-2163-2013
Recall status
Terminated
CDRH record ID
115425
FDA classified
2013-09-06
Recall event number
64147