Z-2163-2020: Panther Fusion Extraction Reagent-S, Ref PRD-04331, IVD, CE, Rx Only consisting of FCR-S Fusion Capture R

Recall record Z-2163-2020 for Hologic, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2163-2020
Company in FDA record
Hologic, Inc
FDA record creation date
May 28, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Reagent component included in flu test kit may cause both high background signal in a true negative sample leading to false positive and/or false negative results. This could result in false positive and/or false negative results on patient samples or failed controls leading to invalid runs. This could result in incorrect results being reported and a delay in treatment.

Source facts

Action
Customers who received product/affected lot were to be contacted via Urgent: Medical Device Recall Notification Letter, dated October 10, 2019 via FedEx overnight. In addition to notifying customers about the recall, the recall notification asked customers to take the following actions: 1) Immediately discontinue use of the lot of Panther Fusion Extraction Reagent-S kits and respective components listed in Table 1. 2) Immediately check inventory and segregate the Panther Fusion Extraction Reagent-S kit lot listed in Table 1. 3) Notified customers of the numbers of kits shipped to customers' site on Customer Response Form. Requested customers complete CRF per the accompanying instructions after segregation of affected product. 4) Requested customers return the completed CRF to Hologic Technical Support, even if they do not have any of the listed kits remaining in their inventory. 5) If customers identified any of the specified kits remaining in inventory (whether unopened or partial kits containing used reagent), requested they contact their technical support using the contact information provided for a replacement. One replacement kit was shipped proactively to minimize downtime due to this issue. 6) After completing the CRF, customers were instructed to destroy the segregated inventory and fax the documentation to the Hologic Technical Support using the fax number +1 858 410 8250 or scan the document(s) and email to Hologic Technical Support at molecularsupport@hologic.com. If customers had any questions or concerns regarding the recall notification, they were to contact Hologic Technical Support at +1 888 484 4747 or +1 858 410 8511 during the hours of 5 am PDT and 5 pm PDT, or by email molecularsupport@hologic.com
Initiated
2019-10-10
Root cause
Nonconforming Material/Component
Distribution
US: AR, CA, FL, ID, IL, IN, LA, NM, NY, PA, TX, UT, VI, WA, WI OUS: AU, CA
Recall number
Z-2163-2020
Recall status
Open, Classified
CDRH record ID
180670
FDA record created
2020-05-28
Recall event number
85346

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗