This notice is being provided in follow up to the field correction initiated by B. Braun Medical Inc. on 26 August, 2011 due to an issue in the Outlook ES Safety Infusion System where the Care Area can inadvertently be exited thereby removing DoseGuard dose limit protection. This occurs when a sequence of HOLD - HOLD key strokes are performed on the pump key panel while the user is in the Dose Dat
Source facts
Action
The firm decided to conduct a field correction to correct the issue and carried it out through routine field services. The firm was notified by FDA that this is a reportable event, so they notified their consignees via letters dated 10/31/2012.
Initiated
2012-10-31
Root cause
Software design
Terminated
2016-02-03
Distribution
Nationwide Distribution.
Recall number
Z-2164-2013
Recall status
Terminated
CDRH record ID
115424
FDA classified
2013-09-06
Recall event number
64146