Maquet/Getinge is initiating a voluntary product removal involving four lot numbers of SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK since during internal pressure (leak) testing of the SENSATION PLUS 8Fr. 50cc, IABC, failures related to the tip seal leak were identified.
Source facts
Action
Getinge notified their customers with a recall letter/return response form via Fed Ex on 4/12/2017.
Getinge notified their customers again through an expansion on 10/16/2017.
Initiated
2017-04-12
Root cause
Device Design
Terminated
2020-01-28
Distribution
Nationwide
Recall number
Z-2164-2017
Recall status
Terminated
CDRH record ID
154647
FDA record created
2018-10-11
Recall event number
76942