Z-2164-2024: SENSE XL TORSO COIL 3.0T. 16-element receive only coil for torso and abdomen imaging in Philips MR Intera

Recall record Z-2164-2024 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2164-2024
Company in FDA record
Philips North America Llc
FDA classification date
Jun 27, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Potential for coils to heat up and harm patients (burn).

Source facts

Action
An URGENT MEDICAL DEVICE RECALL notice, dated 5/31/24, was mailed to consignees. The notice instructs users to follow the provided instructions when operating a MR system with affected coils. These instructions include avoiding First Level Operating Mode/High SAR scans, use dedicated pads and mattresses provided with the coils. avoid positioning the coil closer than 2 inches (5 cm) from the bore, and do not exceed 45 minutes of examination time. An Advisory Notice that summarizes the recommended actions was provided with the recall notification and is to be displayed with affected systems to ensure access by operators. The recall notification is to be circulated among users of affected systems. The completed response form is to be returned to Philips by email at pd.cnr@philips.com. Philips is developing field corrections in response to this recall that include potential software risk control measures to limit scan settings when using the coils, hardware updates, and improving the design of the coil. Philips intends to update consignees on these actions by the end of 2024. Consignees with any questions can contact the Customer Care Solutions center from 8:00 AM to 8:00 PM from Monday through Friday at 1-800-722-9377. An updated letter was mailed to consignees on about 12/11/2024. The contained the same information but provided an update on Philips actions which include: 1. Warning label for coil and pad: Per your response form Philips will contact you to schedule time for a Field Service Engineer (FSE) to visit your site to apply a warning label to the coil and pad starting in January 2025. (Ref 78100607); 2. Instruction For Use (IFU) Update: Including additional warnings and illustrations Q2 2025.
Initiated
2024-06-05
Root cause
Device Design
Distribution
Worldwide distribution: US (Nationwide) and OUS (Foreign): Afghanistan, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Lebanon, Lithuania, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Pakistan, Papua New Guinea, Peru, Philippines, Poland, Portugal, Republic of Korea, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, T¿rkiye, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam.
Recall number
Z-2164-2024
Recall status
Open, Classified
CDRH record ID
208281
FDA classified
2024-06-27
Recall event number
94786

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗