Skip to content

Public FDA record

Z-2165-2020: Omnipod DASH Insulin Management System (mmol/L configuration), Catalog Number PT000011 M/D: INT1D001MM -

Recall record Z-2165-2020 for Insulet Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2165-2020
Company in FDA record
Insulet Corporation
FDA record creation date
May 29, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

After the device has been in use for about 2 months, data processing in the PDM can be slowed such that the Bolus Calculator fails to accurately subtract the correct amout of IOB before suggesting a bolus amount.

Source facts

Action
URGENT: Field Safety Notification letters dated 4/3/20 were sent to customers. How do users prevent this from happening? This issue is entirely preventable (see Figure 1 below) by (1) always entering the Bolus Calculator through the Bolus Button from the Home screen and (2) by never entering the Bolus Calculator through the Menu Icon. Is it safe for users to continue using the Omnipod DASH" PDM? Yes, it is safe for users to continue using the Omnipod DASH" PDM if the steps detailed above are followed to avoid incorrect bolus calculations. Users will not experience the issue if they always use the Bolus Calculator by entering through the Bolus Button on the Home screen. How will Insulet fix the software issue? A software update is being made to fix this issue, along with the issue described in the February 2020 notice and an updated replacement Omnipod DASH" PDM will be provided from August 2020. Once available, we will contact users directly via email or letter to arrange shipping. If you have received an email from Insulet providing you with the information covered in this letter and you have already responded, you do not need to do anything more at this time. If you have not received an email, we would ask that you go our website to confirm receipt of this letter and provide the Serial Number of your Omnipod DASH" PDM to ensure a replacement is reserved for you. If you would like additional information, our Customer Care is available 24/7 on 0800 011 6132 (+44 (0) 203 887 1709 when calling from abroad) or by email omnipod-GB@insulet.com.
Initiated
2020-04-02
Root cause
Device Design
Terminated
2021-09-14
Distribution
The products were distributed to the following foreign countries: Italy, Netherlands, UK.
Recall number
Z-2165-2020
Recall status
Terminated
CDRH record ID
181225
FDA record created
2020-05-29
Recall event number
85552

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗