Skip to content

Public FDA record

Z-2165-2025: Biofinity XR Toric Contact Lens

Recall record Z-2165-2025 for CooperVision, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2165-2025
Company in FDA record
CooperVision, Inc.
FDA classification date
Jul 23, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Lens blisters may have an incomplete or leaking seal which may render them unsterile.

Source facts

Action
An URGENT MEDICAL DEVICE CORRECTION notice dated 6/27/25 was mailed to distributors, retailers, and physicians notifying them of this recall. The notification instructs consignees to inspect their inventory for affected devices and stop any further distribution/sale of affected devices. Practitioners are instructed to share the provided Consumer Communication Letter with any patients who received affected devices and to have patients return any units in their possession for replacement or credit. Consignees are to dispose of any affected product in their inventory or that is returned. Consignees are to complete and return the provided Correction Response Form via fax to 888-385-3217 or by email at info@coopervision.com. Consignees with any complaints can contact the Customer Care Team at 855-526-6737; questions about this recall action can be directed to Customer Service at 800-341-2020.
Initiated
2025-06-16
Root cause
Use error
Distribution
Worldwide distribution - US Nationwide and the countries of Brazil, Canada, Honduras, and Mexico.
Recall number
Z-2165-2025
Recall status
Open, Classified
CDRH record ID
214725
FDA classified
2025-07-23
Recall event number
97173

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗