Open original FDA record ↗Recall record
- FDA identifier
- Z-2166-2011
- Company in FDA record
- Digisonics, Inc
- FDA classification date
- May 10, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Failure of software to update some measurements when tracings or points are amended after measurement calculations have been performed.
Source facts
- Action
- Digisonics, Inc., notified consignees by telephone and e-mail on 10/17/2008, informing them that a problem had been found in the Save Linked Images functionality of DigiView Release 3.6 (and all releases which start with 3.6 such as 3.6.1.). The problem is the failure of the software to update some measurements when tracings or points are amended after measurement calculations have been performed. A utility program, called Linked Image Measurement Verify was being supplied to all sites as a patch to the DigiView program. Release 3.6.6 was to be provided to all customers having previous 3.6 version. Problem was fixed in version 3.6.6 and 3.7.1. If you have any questions, please contact Digisonics Customer Support at (800) 940-3240.
- Initiated
- 2008-11-04
- Root cause
- Device Design
- Terminated
- 2011-05-31
- Distribution
- Worldwide Distribution -- USA including States of : TX, OH, NC, MA, NY, IL, FL, OK, MO, ID, AZ, AR, MI CA, MD, and country of Canada. *** Due to the age of this recall, district is recommending no audit checks***
- Recall number
- Z-2166-2011
- Recall status
- Terminated
- CDRH record ID
- 99270
- FDA classified
- 2011-05-10
- Recall event number
- 58400