Skip to content

Public FDA record

Z-2166-2013: Outlook Safety Infusion System Burette Set. For intravenous fluid administration.

Recall record Z-2166-2013 for B. Braun Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2166-2013
Company in FDA record
B. Braun Medical, Inc.
FDA classification date
Sep 6, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Reports were received of an incorrectly assembled Outlook Safety Infusion System Burette Set (Product Code 375038), and the investigation determined that one of the ULTRASITE injection sites and corresponding low pressure backcheck valve was assembled upside down.

Source facts

Action
On August 4, 2013, a letter was sent to direct accounts asking them to check their inventory for the affected product. These customers were also instructed to notify their customers if the product was further distributed and to return all affected product and/or the product removal acknowledgement form.
Initiated
2013-08-05
Root cause
Process control
Terminated
2014-09-03
Distribution
Distributed in the states of FL, NY, NJ, and TX.
Recall number
Z-2166-2013
Recall status
Terminated
CDRH record ID
121164
FDA classified
2013-09-06
Recall event number
65993

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗