Z-2166-2021: basixALPHA Inflation Device 30 atm/bar 20 mL - Product Usage: Used to inflate and deflate an angioplasty

Recall record Z-2166-2021 for Merit Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2166-2021
Company in FDA record
Merit Medical Systems, Inc.
FDA record creation date
Jul 29, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a design issue, inflation syringe handle separating from the inflation device when handle is withdrawn.

Source facts

Action
On 06/08/2021, the firm sent an "URGENT PRODUCT RECALL NOTICE" initially via an email communication and then followed up with a hardcopy letter via FedEx overnight on 06/10/2021. The communication and letter informed customers that due to a design defect the handle of the inflation device may separate from the syringe during use, which may result in a delay in procedure. Customers are instructed to: 1. Please immediately determine if any of the devices identified in the attached Customer Response Form (CRF) are within their facility, quarantine them, and discontinue use and distribution. 2. Ensure that applicable personnel within their organization are made aware of this recall. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, ensure this notice is immediately shared with them and note the quantity distributed on the CRF. Additional distribution details may be required by health authorities. 4. Fill out, scan, and email the completed CRF to Customer Service at response@merit.com within 10 business days. All affected product shipped must be accounted for on the CRF. 5. Immediately return all affected lots in their possession to the Recalling Firm, per the instructions in the attached CRF. For any question contact the Recalling Firm's Representative or Customer Service at 801-208-4381 Monday thru Friday 6:00 am to 6 pm MDT.
Initiated
2021-06-10
Root cause
Device Design
Terminated
2024-05-15
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of AL, AZ, CA, CO, DE, FL, GA, ID, IL IN, KS, LA, MI, MN, MO, NC, NE, NJ, NV, NY, OH, OR, PA, SC, TN, TX, UT, WA, WI, WV, and WY. The countries of Australia, Austria, China, Denmark, Finland, France, Germany, Israel, Japan, Kenya, Kyrgyzstan, Netherlands, New Zealand, Norway, Qatar, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates and United Kingdom.
Recall number
Z-2166-2021
Recall status
Terminated
CDRH record ID
188062
FDA record created
2021-07-29
Recall event number
88191

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗