Open original FDA record ↗Recall record
- FDA identifier
- Z-2166-2024
- Company in FDA record
- Horiba Instruments Incorporated
- FDA classification date
- Jul 1, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
HORIBA custom configured fluorescence instrument has a laser safety defect. This custom configuration of the modular Instrument removed the primary light source and the excitation monochromator. A plastic cap was installed to plug the optical port, where a more durable beam stop should have been used. The defect may allow the class 4 laser beam to burn (or transmit) through the plastic cap to exit the protective enclosure.
Source facts
- Action
- Horiba plans to issue an IMPORTANT ELECTRONIC PRODUCT RADIATION WARNING letter to notify affected customers. The notification will Identify the affected product involved, description of the defect, safety hazard imposed: and corrective actions, For questions, call (732) 623-8146.
- Initiated
- 2024-03-28
- Root cause
- Radiation Control for Health and Safety Act
- Distribution
- US Nationwide and Worldwide Distribution
- Recall number
- Z-2166-2024
- Recall status
- Open, Classified
- CDRH record ID
- 208478
- FDA classified
- 2024-07-01
- Recall event number
- 94839