Z-2166-2026: Thermedx FluidSmart Urology Tube Set REF LL0006 UDI-DI code: 00860936000365 The FluidSmart Tube Sets are

Recall record Z-2166-2026 for Stryker Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2166-2026
Company in FDA record
Stryker Corporation
FDA classification date
May 12, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to nonconforming products being inadvertently distributed.

Source facts

Action
On 04/09/2026, the firm sent via USPS Certified Mail an "URGENT: MEDICAL DEVICE RECALL" Letter to customers that Stryker identified that certain lots of FluidSmart tube sets were manufactured with an incorrect pressure relief valve specification. This defect may prevent the pressure relief valve from activating at the intended pressure in the event of a capital device over pressure failure. One lot (Lot B000576601, quantity 10) impacted by this nonconformance was inadvertently shipped to a customer. Customers are instructed to: 1. Inform individuals within their organization who need to be aware of this action. 2. Check internal inventory to locate the product listed in the product affected grid. 3. If affected product is found, stop use immediately, segregate the product and call Stryker Customer Service at 1-800-624-4422 (Option 3) or email endocustomersupport@stryker.com to arrange for product return and issuance of credit or replacement (upon availability). a. Remove the affected products, quarantine, and discontinue use of the identified product. b. Please complete Business Reply Form on Attachment A on Page 2 and return it to EndoRecall@stryker.com. 4. If affected product is NOT found: a. Please complete Business Reply Form on Attachment A on Page 2 and return it to EndoRecall@stryker.com. 5. Maintain awareness of this communication internally until all required actions have been completed within your facility. Response is required. For questions - EndoRecall@stryker.com or 1-800-624-4422 (Option 3).
Initiated
2026-04-09
Root cause
Process change control
Distribution
U.S. Nationwide distribution in the state of GA.
Recall number
Z-2166-2026
Recall status
Open, Classified
CDRH record ID
219497
FDA classified
2026-05-12
Recall event number
98696

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗