Reason stated in the source
Software: two displayed respiratory parameters are incorrect due to errors in the system software: MValv (alveolar minute ventilation) and Vt/kg (tidal volume/patient weight). In both instances, the displayed parameters are higher than actual.
Source facts
- Action
- Respironics issued Field Action Notification Letter via Fed'x on 1/4/11, notifying customers of the issue and to instruct them on conditions for continued use and return of the device. The device will be updated with revised software to correct the calculation errors. Any questions or need additional information about this recall action, Contact Respironics US Customer Service at 1-800-345-6443 for customers residing in the USA or +011 (203)697-6488 for International customers.
- Initiated
- 2011-01-04
- Root cause
- Software design
- Terminated
- 2013-03-07
- Distribution
- Nationwide Foreign: Belgium, China, Columbia, Egypt, England, India, Italy, Saudi Arabia, and Spain
- Recall number
- Z-2167-2011
- Recall status
- Terminated
- CDRH record ID
- 96739
- FDA classified
- 2011-05-10
- Recall event number
- 57594