Z-2168-2008: Suction Jar Lid for Medela Vario 18 Vacuum Pump used with the Bravo pH Monitoring System Medtronic Part N

Recall record Z-2168-2008 for Medtronic Neuromodulation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2168-2008
Company in FDA record
Medtronic Neuromodulation
FDA classification date
Sep 4, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The Medela Vario 18 Vacuum Pump used with the Medtronic Bravo pH Monitoring System. The Vacuum Pump is used to provide the vacuum pressure during Capsule attachment. Some Medela Vario 18 Vacuum Pumps were provided with an incorrect suction Jar Lid that did not have the correct conical connection for the Vacuum Tube.

Source facts

Action
Affected customers contacted by letter "Medtronic Urgent: Medical device Recall April 2008" that indicates that some Medela Vario 18 Vacuum Pumps were provided with an incorrect Suction Jar Lid that did not have the correct Conical Connection for the Vacuum Tube and the method to obtain a replacement part if needed. The letter requires customers to locate the Vacuum Pump serial number and compare it against the Serial Number on the Reply Form provided. Next customers are to review the Suction Jar Lid and check the part number. "Reply form and Instruction for Receiving Replacement Lid " included. Contact Medtronic 1-800-707-0933 for assistance.
Initiated
2008-04-25
Root cause
Employee error
Terminated
2012-05-15
Distribution
USA including states of OH, DE, CT, KY, AR, NY, MA, CA, GA, NC, MI, WI, TX, PA, MD, and country of CANADA.
Recall number
Z-2168-2008
Recall status
Terminated
CDRH record ID
71725
FDA classified
2008-09-04
Recall event number
48612

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗