Skip to content

Public FDA record

Z-2168-2014: Nanosphere Verigene Each CDF test consists of a Verigene CDF Nucleic Acid Test Kit and a Verigene CDF

Recall record Z-2168-2014 for Nanosphere, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2168-2014
Company in FDA record
Nanosphere, Inc.
FDA classification date
Aug 8, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Nanosphere has received report of an increased rate of "No Call - INT CTL 2" results for the Internal Control 2 (INT CTL 2) when negative samples are tested with certain lots of CDF Amplification Trays (20-011-022), which are contained in the CDF Amplification Reagent Kit (20-012-022); specifically for lots 061913022D and 071813022E.

Source facts

Action
Nanosphere sent an Field Notification Letter dated October 16, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letters included instructions for customers to: 1) discontinue use of the affected product, 2) discard any of the affected product, and 3) complete the attached form and return it to Nanosphere. Customers with questions can contact Nanosphere at 847-400-9000 or 888-837-4436 x 2.
Initiated
2013-10-16
Root cause
Nonconforming Material/Component
Terminated
2015-05-22
Distribution
US Distribution including the states of CA and WA
Recall number
Z-2168-2014
Recall status
Terminated
CDRH record ID
128487
FDA classified
2014-08-08
Recall event number
68708

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗