Z-2168-2025: Medtronic CareLink SmartSync Patient Connector, Model Number 24967 with the following software components

Recall record Z-2168-2025 for Medtronic, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2168-2025
Company in FDA record
Medtronic, Inc.
FDA classification date
Jul 23, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for the user to abort a therapy, thereby limiting the user's ability to cancel a high voltage therapy delivery.

Source facts

Action
An URGENT MEDICAL DEVICE CORRECTION NOTIFICATION letter dated 6/18/25 was sent to customers. Customer Actions: "Update SmartSync (to version 4.2.3 or higher). Refer to release notes provided by your local Medtronicrepresentative for update instructions. If needed, Medtronic representatives are available to work with you toinstall or update the SmartSync application(s) on your tablet. See Appendix for instructions on how to verify thesoftware update is complete. "Sign and return the enclosed confirmation form and share this notice with those who need to be aware withinyour organization or with any organization where SmartSync may be in use. Questions regarding this informationshould be directed to the local Medtronic Representative or Medtronic Programmer Technical Services at 800-638-1991, Option 1.
Initiated
2025-06-18
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution.
Recall number
Z-2168-2025
Recall status
Open, Classified
CDRH record ID
214626
FDA classified
2025-07-23
Recall event number
97130

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗