Open original FDA record ↗Recall record
- FDA identifier
- Z-2169-2010
- Company in FDA record
- Hill-Rom, Inc.
- FDA classification date
- Aug 3, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The flexing of the junction of the sleep deck's head and mid sections can create excess pressure on the pivot pin E-ring, which could cause the E-ring to come off.
Source facts
- Action
- The firm sent an Urgent Field Safety Notice to its customers dated 9/16/2009. The letter identified the affected product, explained the reason for the recall, actions to be taken, and a contact reference. Customers are to locate the affected product, examine the pivot points and ensure the E-ring is present. If the E-ring is missing, then the affected bed is to be taken out of service. Customers' distributors or agents will contact them with a permanent fix. Questions should be directed towards Hill-Rom Technical Support.
- Initiated
- 2009-09-16
- Root cause
- Device Design
- Terminated
- 2012-03-16
- Distribution
- Worldwide Distribution -- Argentina, Bahrain, Bhutan, Brazil, Bulgaria, Chili, China, Columbia, Costa Rica, Ecuador, Egypt, Eritrea, France, Hong Kong, India, Indonesia, Iran, Iraq, Jordan, Kenya, Kuwait, Lebanon, Libya, Malaysia, Mexico, Morocco, Oman, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Puerto Rico, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Syria, Taiwan, Thailand, Trinidad & Tobago, Turkey, United Arab Emirates, United States, Uruguay, Venezuela, and Vietnam.
- Recall number
- Z-2169-2010
- Recall status
- Terminated
- CDRH record ID
- 92433
- FDA classified
- 2010-08-03
- Recall event number
- 56018