Skip to content

Public FDA record

Z-2170-2021: Nail Insertion Sleeve, elastic T2 Tibia SPI 8-13 (SPI Elast. Sleeve 8-13), Catalog Number 18061407S

Recall record Z-2170-2021 for Stryker GmbH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2170-2021
Company in FDA record
Stryker GmbH
FDA record creation date
Jul 30, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The devices have a different inner diameter than the diameter specified on the outer box label.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL notice dated 5/27/21 was sent to customers. Actions to be Taken by the Customer/User: Our records indicate that you may have received one or more of the affected products. It is Strykers responsibility as the manufacturer to ensure that customers who may have received these affected products also receive this important communication. We therefore request that you read this notice carefully and complete the following actions. 1. Please inform users of this Urgent Medical Device Recall and forward this notice to all individuals who need to be aware within your organization. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list is at your facility. 3. Quarantine and discontinue use of the recalled devices identified in the affected product list. Return the devices back to Stryker. 4. Under 21 CFR 803, manufactures are also required to report any serious injuries where a product has contributed or may have contributed to the event. Please keep Stryker informed of any adverse events associated with this product by emailing soprodexpreports@stryker.com. 5. A response is required, even though you may not have any physical inventory on site anymore. 6. Hospitals/Branches/Agencies: Complete and sign the enclosed Urgent Medical Device Recall Business Reply Form and fax a copy to 1-888-879-4609 or email to StrykerOrtho8434@Stericycle.com. 7. Hospitals Only: Please contact your Local Sales Office or your Stryker Sales Representative directly for product returns and inventory questions. 8. Branches/Agencies Only: Return all affected devices available at your location to the following address. Stryker Orthopaedics/PFA Product Returns Attn: Distribution Inventory Team 325 Corporate Drive Dock M-East Mahwah, NJ 07430 Ref. PFA 2644769 Please assist us in meeting our regulatory obligation by faxing back the attached Urgent Medical Device Recall Business
Initiated
2021-05-27
Root cause
Process control
Distribution
US Nationwide Distribution
Recall number
Z-2170-2021
Recall status
Open, Classified
CDRH record ID
188204
FDA record created
2021-07-30
Recall event number
88240

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗