Open original FDA record ↗Recall record
- FDA identifier
- Z-2171-2010
- Company in FDA record
- Abbott Laboratories
- FDA classification date
- Aug 3, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
IMx Sirolimus reagent lot 802873106 has exhibited an increased frequency of calibration errors.
Source facts
- Action
- Abbott sent a "Product Recall" letter dated April 30, 2010, to all customers receiving the affected lot of reagent. The accounts were informed of the increased frequency of calibration errors, including: 157 Check 1 Out of Range, 161 Check 5 Out of Range, and 173 Mode 1 deviation Factor High. The accounts were requested to determine if they are currently using or have inventory of lot 802873106, discontinue use of and destroy any remaining inventory of the lot, and retain a copy of the letter for their laboratory records. Customers requiring additional information were directed to contact Abbott Customer Service at 1-877-422-2688. Customers outside of the U.S., please contact your local Customer Service Representative.
- Initiated
- 2010-04-30
- Root cause
- Other
- Terminated
- 2011-05-13
- Distribution
- Worldwide distribution: USA and countries including Puerto Rico, Argentina, Austria, Bahrain, Brazil, Canada, Chile, China, Croatia, Cyprus, Egypt, France, Germany, Greece, Hungary, India, Israel, Italy, Japan, Jordan, Lebanon, Lithuania, Netherlands, Norway, Philippines, Poland, Russia, Panama, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE and Venezuela.
- Recall number
- Z-2171-2010
- Recall status
- Terminated
- CDRH record ID
- 91137
- FDA classified
- 2010-08-03
- Recall event number
- 55557