Open original FDA record ↗Recall record
- FDA identifier
- Z-2171-2012
- Company in FDA record
- Abbott Point Of Care Inc.
- FDA classification date
- Aug 9, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
The frequency of suppressed results for i-Stat cTnl, BNP and CK-MP cartridges is affected by atmospheric pressure.
Source facts
- Action
- Abbott Point of Care (APOC) sent an URGENT RECALL NOTICE dated February 2012 to all affected customers via e-mail and also contacted by phone. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised that if their facility was performing testing at more than approximately 7500 feet above sea level, they should have an alternate test method available should they experience an issue. If they have forwarded any cartridges to another facility, they should provide a copy of the Urgent Recall Notice to them. For questions customers were instructed to call 800-366-8020, Option 1 or contact their Abbott Point of Care representative. For questions regarding this recall call 609-454-9000.
- Initiated
- 2012-02-01
- Root cause
- Device Design
- Terminated
- 2013-02-08
- Distribution
- Worldwide Distribution - USA including AZ and CO and internationally to Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Chile, China, Colombia, Costa Rica, Dominican Republic, El Salvador, Finland, France, Georgia, Germany, Greece, Honduras, Hong Kong, India, Ireland, Italy, Korea, Kuwait, Mauritius, Mexico, Netherland (Antilles), Nigeria, Norway, Oman, Paraguay, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, R.O.C., Thailand, UAE, United Kingdom, and Yemen.
- Recall number
- Z-2171-2012
- Recall status
- Terminated
- CDRH record ID
- 110870
- FDA classified
- 2012-08-09
- Recall event number
- 62533