Z-2171-2024: CARESCAPE B650: Model Numbers: a) MBB313, Model Numbers: 1) 2095801-001, 2) 2095801-001-01018416,

Recall record Z-2171-2024 for GE Healthcare Finland Oy. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2171-2024
Company in FDA record
GE Healthcare Finland Oy
FDA classification date
Jun 21, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

GE HealthCare has become aware of the potential for loss of monitoring on CARESCAPE Canvas 1000, CARESCAPE Canvas Smart Display, and certain versions of CARESCAPE B850 and CARESCAPE B650 monitors if, following the replacement of the CPU timekeeper battery, a mains power loss occurs during active monitoring. In the unlikely event that this occurs, it can result in the displays or monitors not turning on, after the mains power is restored. This can result in a delay in recognizing changes in patient condition until a replacement monitor is connected.

Source facts

Action
GE HealthCare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 04/05/2024 via *****. The notice explained the problem with the device, potential risk, and requested the following actions to be taken by Customer/User: 1. You can continue to use your monitor if the CPU timekeeper battery has not been replaced. If the monitor has had a CPU timekeeper battery replaced or if you are unsure whether the CPU timekeeper battery was replaced, reinstall the host software using the software media that is included as part of the equipment. 2. In the event the monitor does not start up following a mains power loss during active monitoring, continue patient assessment with other methods until a replacement monitor or display has been connected. 3. Update any facility planned maintenance schedules to remove CPU timekeeper battery replacement. Only change the battery if "Service Monitor Error Code 0xHOST1100" error message occurs. After changing the battery, reinstall the host software using the host software media that is included as part of the equipment. 4. Download the full User manual documentation set which contains the revised Service and Supplemental Information Manual(s) from the GE HealthCare Customer Documentation Portal. See Appendix for a list of updated and impacted part numbers. https://www.gehealthcare.com/support/manuals If you have difficulty accessing the documentation, please contact GE HealthCare for assistance. 5. Discontinue usage and destroy and/or delete all impacted versions of the Service Manuals, Supplemental Information Manuals and delivery media. 6. Complete and return the attached Acknowledgement Form to FMI.36161@gehealthcare.com.
Initiated
2024-04-05
Root cause
Software design
Distribution
Worldwide distribution.
Recall number
Z-2171-2024
Recall status
Open, Classified
CDRH record ID
207315
FDA classified
2024-06-21
Recall event number
94494

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗