Skip to content

Public FDA record

Z-2171-2026: ZYMUTEST HIA IgG, in vitro diagnostic ELISA kit, Ref. RK040A

Recall record Z-2171-2026 for Aniara Diagnostica LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2171-2026
Company in FDA record
Aniara Diagnostica LLC
FDA classification date
May 12, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Kits from lot FD1265 had been packaged on 17 October 2024 and put on the market with an outdated flyer (8 October 2024), containing incorrect values for the positive control.

Source facts

Action
A Field Corrective Action notification letter dated 3/30/26 was sent to customers. Actions implemented by HYPHEN BioMed: -Confirmation of the existence of a flyer discrepancy affecting 95 kits from lot FD1265. - Request for quarantine of the affected kits in US agent warehouse. -Coordination with US agent for notification of the impacted end users. -investigation to define the root cause and to establish corrective and preventive actions. Actions to be implemented by the end user: 1. Check the TAV flyer included in their kit. - If the flyer matches the correct version (dated 18 October 2024) - No action is required. - If the flyer corresponds to the incorrect version (dated I October 2024) - proceed to step 2 2. Destroy the incorrect TAV flyer. The outdated flyer included in the kit must be immediately removed and destroyed, as well as any paper or electronic copy that could have been done in end-user facility, to prevent further use. 3. Replace the incorrect flyer with the corrected version. Use the corrected TAV flyer provided with this notice for all future testing 4. Return the completed and signed form to your local distributor before April 15th, 2026. For any question or information regarding this notification, please contact vigilance team, by email at VigilanceHBM@hyphen-biomed.com.
Initiated
2026-04-01
Root cause
Labeling Change Control
Distribution
US Nationwide distribution in the states of CA, GA, MN, MO, NC, NM, NY, VA, WA, and WI
Recall number
Z-2171-2026
Recall status
Open, Classified
CDRH record ID
219697
FDA classified
2026-05-12
Recall event number
98754

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗