Open original FDA record ↗Recall record
- FDA identifier
- Z-2172-2017
- Company in FDA record
- Laserex Systems Inc.
- FDA classification date
- Jun 6, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
It was discovered the unit produced a laser emission without pressing the fire button when the slit-lamp was driven to its lowest position prior to use on any patient.
Source facts
- Action
- Ellex Medical shall, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard. 1. Installation of a spacer/collar to prevent the slit lamp from being lowered to the point where the cable can be crushed, 2. Improvement of cable management, 3. The corrections will be conducted at no cost to the purchaser, and 4. The corrective action will be completed by June 30, 2017. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. For further questions, please call (800) 824-7444.
- Initiated
- 2017-04-04
- Root cause
- Device Design
- Terminated
- 2017-07-31
- Distribution
- US Distribution
- Recall number
- Z-2172-2017
- Recall status
- Terminated
- CDRH record ID
- 155147
- FDA classified
- 2017-06-06
- Recall event number
- 77170